Lukas T. Rotkopf, Marco Schlimbach, Kevin Kröninger, Stefan O. Schoenberg, Heinz-Peter Schlemmer, Jens Kleesiek, Moritz Rempe

VarNet-Seg jointly trains MRI reconstruction and lesion segmentation for prostate MRI with retrospective undersampling. It combines learned reconstruction, data-consistency updates, and a 3D segmenter trained end to end. Dice, a measure of mask overlap, was 0.222 for combined PIRADS 4 and 5 lesions at simulated eight-fold undersampling, 0.078 above fixed SENSE-Seg; patient-level AUROC fell to 0.633 from 0.750 at full sampling.